Electrical Requirements

 

Power Source

An independent 20-A line (120-V systems) should be provided for the system (10 A for 220-V systems), including the computer. The power source should not be associated with heavy-duty equipment such as large motors, or with possible sources of high-frequency interference such as photocopying systems, discharge lamps, or radio transmitters.

Line Voltage

PerkinElmer analyzers and their associated instruments are designed to operate within a line voltage range of 10% of the nameplate voltage or other such voltage selected at installation to suit the particular country or region. (The range for 240-V systems is +6%, –10%.) The supply must be smooth, clean, and free of transient voltages over 40 V.
The frequency range is +/-1% for 50-Hz and 60-Hz systems.

Instrument Maximum Power Requirements

The maximum power requirements for the analyzers are as follows:

Diamond DSC : 300 VA

Pyris 1 DSC: 300 VA

CryoFill LN2 Control Box: 50 VA

Pyris 6 DSC: 225 VA   

Pyris 1 TGA : 2270 VA   

Pyris 6 TGA: 250 VA   

Diamond DMA 800VA