Electrical Requirements
Power Source
An independent 20-A line (120-V systems) should be provided for the system (10 A for 220-V systems), including the computer. The power source should not be associated with heavy-duty equipment such as large motors, or with possible sources of high-frequency interference such as photocopying systems, discharge lamps, or radio transmitters.
Line Voltage
PerkinElmer analyzers and their associated instruments are
designed to operate within a line voltage range of 10% of the nameplate voltage
or other such voltage selected at installation to suit the particular country
or region. (The range for 240-V systems is +6%, 10%.) The supply must
be smooth, clean, and free of transient voltages over 40 V.
The frequency range is +/-1% for 50-Hz and 60-Hz systems.
Instrument Maximum Power Requirements
The maximum power requirements for the analyzers are as follows:
Diamond DSC : 300 VA
Pyris 1 DSC: 300 VA
CryoFill LN2 Control Box: 50 VA
Pyris 6 DSC: 225 VA
Pyris 1 TGA : 2270 VA
Pyris 6 TGA: 250 VA
Diamond DMA 800VA