Environmental Requirements

 

You must provide the following laboratory conditions for your Thermal Analysis System:

 

A clean area free of strong magnetic fields.

An adequate and stable power source for all system components.

The area must have a relative humidity of less than 50%. The temperature of the area should be between 10°C and 40°C (39°F and 104°F).

 

Place the system components in an area that is not in direct sunlight or direct contact with heating and cooling ducts or units.

The instruments are for indoor use only.

The altitude limitation is 2000 m.

The overvoltage category for all instruments is Category II.

The pollution degree is 2 for all instruments.

If the equipment is used in a manner other than instructed, the protection provided may be impaired